Remote control commands
R&S
®
RTO6
2431
User Manual 1801.6687.02 ─ 05
ACQuire:POINts:DVALue?
Returns the current digital record length used by each digital channel.
Return values:
<DigReclgth>
Range:
1000 to 200E6
Increment: 2
*RST:
1000
Default unit: pts
Usage:
Query only
Asynchronous command
24.18.4
Trigger settings for digital signals and parallel buses
In all
TRIGger<m>:PARallel
commands, the trigger suffix <m> is always 1 and can
be omitted. It selects the trigger event: Only the A-trigger is available for triggering on
digital signals and parallel buses.
.............................................................................................2431
............................................................................................2436
............................................................................................2440
24.18.4.1
General commands
.....................................................................................2431
.........................................................................................2432
TRIGger<m>:PARallel:SETHold:CSOurce[:VALue]
.......................................................... 2433
TRIGger<m>:PARallel:STATe:CSOurce:VALue
................................................................2433
TRIGger<m>:PARallel:SPATtern:CSOurce[:VALue]
......................................................... 2433
TRIGger<m>:PARallel:EDGE:EXPRession[:DEFine]
........................................................2433
TRIGger<m>:PARallel:WIDTh:EXPRession[:DEFine]
.......................................................2433
TRIGger<m>:PARallel:TIMeout:EXPRession[:DEFine]
.....................................................2433
TRIGger<m>:PARallel:STATe:EXPRession[:DEFine]
........................................................2433
TRIGger<m>:PARallel:PATTern:EXPRession[:DEFine]
.....................................................2433
TRIGger<m>:PARallel:SPATtern:EXPRession[:DEFine]
................................................... 2433
TRIGger<m>:SOURce[:SELect]
<SourceDetailed>
Selects the source of the trigger signal.
Mixed signal option (MSO, R&S
RTO6-B1)