Remote control commands
R&S
®
RTO6
2440
User Manual 1801.6687.02 ─ 05
TRIGger<m>:PARallel:PATTern:WIDTh:DELTa
<WidthDelta>
Defines a range around the width value set with
.
Parameters:
<WidthDelta>
Range:
0 to 432
Increment: 500E-12
*RST:
0
Default unit: s
Usage:
Asynchronous command
24.18.4.8
Serial pattern trigger
See also:
●
TRIGger<m>:PARallel:SPATtern:CSOurce[:VALue]
●
TRIGger<m>:PARallel:SPATtern:EXPRession[:DEFine]
TRIGger<m>:PARallel:SPATtern:CSOurce:EDGE
<ClockEdge>
Sets the edge of the clock signal. The bit value is determined at the crossing of the
clock edge and the logical threshold.
Parameters:
<ClockEdge>
POSitive | NEGative | EITHer
*RST:
POSitive
Usage:
Asynchronous command
TRIGger<m>:PARallel:SPATtern:PATTern
<Pattern>
Defines the serial bit string on which to trigger.
Parameters:
<Pattern>
Numeric or string pattern, see
on page 1286. The bit value X (don't care) is not
allowed.
Usage:
Asynchronous command
24.18.5
MSO data
To export data of digital channels and parallel buses to file, use the following com-
mands:
●
●
●
Mixed signal option (MSO, R&S
RTO6-B1)