Functional Check
R&S
®
RT-ZS60
41
User Manual 1418.7342.02 ─ 04
6
Functional Check
The functional check confirms the basic operation of the R&S
RT-ZS60 active
voltage probe. The functional check is not suitable for verifying compliance with
the probe specifications.
1. Connect the R&S
RT-ZS60 probe to a Rohde
&
Schwarz oscilloscope as
described in
Chapter 2.1, "Connecting the Probe to the Oscilloscope"
,
on page 15.
2. Connect the signal pin to the square wave output of the oscilloscope.
3. Connect the ground pin to the probe ground connector of the oscilloscope.
4. Press the [Preset] key and then the [Autoset] key on the oscilloscope.
A square wave with 1
V amplitude between 0
V and 1
V is displayed on the
display.