Remote control
R&S
®
QAR50
58
User Manual 1179.4992.02 ─ 02
During a remote session, the user interface is unavailable. To access the user interface
again, you have to return to local mode via the user interface or end the remote ses-
sion with
Example:
//Initiate remote session
>R
//Send commands
MEAS:START
MEAS:SAVE 'A great measurement'
Usage:
Event
MEASurement:EVALuation:WINDow
<XSize>, <YSize>
Defines the size of the evaluation window.
Setting parameters:
<XSize>
Length of the horizontal side of the evaluation window in mm.
Range:
1 to 100
<YSize>
Length of the vertical side of the evaluation window in mm.
Range:
1 to 100
Example:
See
Chapter 12.3, "Remote control examples"
Usage:
Setting only
MEASurement:LOAD
<Directory>
Restores previously recorded measurement data.
The restored data depends on what was saved (
).
Setting parameters:
<Directory>
String containing the path to the directory that contains the data-
set.
Example:
//Restore measurement data
MEAS:LOAD 'c:
\qar50\measurements\20211111-111111_measurement0001'
Usage:
Setting only
MEASurement:NORMalize:REFLection:REQuired?
Queries if the reflection measurement has been normalized.
Remote control commands