MSRA basics
R&S
®
FSWP-B1
30
User Manual 1177.5862.02 ─ 09
Limitations
However, since the data in the I/Q Analyzer secondary application is captured by the
Primary, independently of the specific time or frequency measurement requirements
concerning the RBW, filter type and number of sweep points in the secondary applica-
tion, some restrictions may apply to these measurements in the I/Q Analyzer. If not
enough samples are available in the captured and converted I/Q data, for example, an
error message is displayed in the secondary application.
The
maximum span
for a frequency sweep on I/Q-based data corresponds to the
maximum I/Q bandwidth.
The
maximum resolution bandwidth (RBW)
is 1
MHz.
Furthermore, the following
functions
are not available for time and frequency domain
measurements in multistandard mode:
●
Marker demodulation
●
Frequency counter marker
●
Gated measurement
●
Video trigger