Reference: format description for I/Q data files
R&S
®
FSWP-B1
118
User Manual 1177.5862.02 ─ 09
Annex
A
Reference: format description for I/Q data
files
This section describes how I/Q data is transferred to the memory during remote control
(see
command).
For details on the format of the individual values, see
Chapter B, "Formats for returned
values: ASCII format and binary format"
For details on the format of I/Q export files (using the "I/Q Export" function), see the
R&S
FSWP User Manual.
Figure A-1: I/Q data formats
Note
: 512k corresponds to 524288 samples
For maximum performance, the formats "Compatible" or "IQPair" should be used. Fur-
thermore, for large amounts of data, the data should be in binary format to improve
performance.
In binary format, the number of I- and Q-data can be calculated as follows: