Adjustments
4
R30
2/A
- Adjustment
s
Ralco
srl
Adjustment Procedures
• Remove the part of the cover to access the point of
adjustment. See Collimator
C
OVER
R
EMOVAL
• The line is to fall on a perpendicular corss-line on
the antidust plastic panel near the collimator con-
trols.
• Adjust the position of the line by rotating or moving
the base of the laser system.
• To rotate the laser system, loosen the Allen screw
on the metal strip holding the laser. See
Laser
Adjustment.
• Tighten the Allen screw when the laser beam falls
on or is parallel to the bisector line drawn on the
anti-dust panel.
• Shift the laser system by loosening the two screws holding the laser system base to the
beam limiting device front plate.
• Move the base until the laser beam falls over the perpendicular bisector line on the anti-
dust panel.
• Tighten the two screws.
NB:
DO
NOT
APPLY
EXCESSIVE
FORCE
TO
THE
SCREW
. T
HE
LASER
SHELL
IS
PLASTIC
AND
EXCESSIVE
PRESSURE
COULD
CRACK
THE
PLASTIC
AND
POSSIBLY
SHORT
-
CIRCUIT
THE
LASER
.
Summary of Contents for R 302F/A