MTR302/A - R302F/A - R302L/A DHHS
MANUALE ISTRUZIONI / INSTRUCTION MANUAL
Modello/Model R302/A
Compliance Verification
9
R302/A
- Compliance V
erification
Ralco
srl
Electromagnetic Immunity for Non Life-Supporting System
Table 2:
T
HE
R302/A
COLLIMATOR
IS
INTENDED
FOR
USE
IN
THE
ELECTROMAGNETIC
ENVIRONMENT
SPECIFIED
BELOW
. T
HE
CUSTOMER
OR
THE
USER
OF
THE
COLLIMATOR
SHOULD
ASSURE
THAT
IT
IS
USED
IN
SUCH
AN
ENVIRONMENT
.
Immunity Test
IEC 60601 Test
Level
Complian
ce Level
Electromagnetic
Environment
Portable and mobile RF
communications equipment
should be used no closer to
any part of the R302/A,
including cables, then the
recommended separation
distance calculated from
the equation applicable to
the frequency of the
transmitter.
Recommended
Separation Distance
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2,5 GHz
3 V/m
d
=1.2
√
P
80 MHz to
800MHz
d
=2.3
√
P
800
MHz to 2.5GHz
Conducted RF
IEC 61000-4-6
3 V/m
150 kHz to 80 GHz
3 V
d= 1.2 x
√
P
Where
P
is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and
d
is the
recommended separation
distance in metres (m).
Summary of Contents for R 302F/A