• Check to see that the distance between the collimator housing and the mounting flange is
equal in all directions and that the collimator face is parallel to the axis of the table.
• Loosen the screws and adjust if required.
• Tube/collimator mounting flange
IF
THE
COLLIMATOR
IS
TO
BE
MOUNTED
ON
A
ROTATING
FLANGE
,
USE
A
TIGHTENING
FORCE
BETWEEN
MIN
. 0,50
NM
AND
MAX
. 0.75
NM
.