EN 60601-1 par. 5
• Protection against electric hazards: "Class I" equipment.
• Protection against direct and indirect contacts: Type B equipment with applied parts.
• Protection against water see page: "Common equipment".
• Safety of operation in the presence of inflammable anaesthetics with air or oxygen or
nitrous oxide: Equipment not suited to application in the presence of inflammable anaes-
thetic mixtures containing air or oxygen or nitrous oxide.
• Operation conditions: Equipment for continuous operation at intermittent loads - See Chap-
ter -
OPERATION INSTRUCTIONS
.
• Should label data on the collimator not correspond to the specifications herein, inform
Ralco of the non conformity.
• Verifications of the specifications are to be performed according to the indicated equipment
standards.
Operation Environment
• Ambient temperature = from 10°C to 40°C
• Relative Humidity = from 30% to 75%
• Atm. Press. = from 700 to 1060 hPa
Storage
• Temperature from -40°C to 70°C
• Humidity from 10% to 95%
• Atm. Pressure from 500 to 1600 hPa
The device may be stored in its original packing for some time before it is installed. However,
even if each metal part has been accurately protected with specific varnish or superficial
oxidation treatment, the device has not been designed for outdoor use or storage. The device
is sensitive to humidity; avoid keeping it outdoors or using it in unsuitable conditions.
EN 60601-1-3 para. 29.205 - Focus-Skin Distance
Paragraphs 29.205.1 - 29.205.3 of the above standard deal with preventing the use of
inappropriate small focus/skin distances in order that the dose equivalent to the patient be kept
as low as reasonably achievable.
A focus-skin spacer for each collimator type is available on request.
29.205.1 - Radiation Equipment for Fluoroscopy
X-ray devices for fluoroscopy must feature characteristics that, during fluoroscopy, will prevent
focus-skin distances less than:
• 20 cm if the x-ray device is specifically for fluoroscopy.
• 30 cm for other applications.
Conformity is measured and checked visually. The x-ray system must comply with the norm;
this means that the operator is responsible for the verification. A focus/skin spacer device is an
option made available by Ralco.
GUIDES
FOR
ACCESSORIES
(
IF
ASSEMBLED
):
PRIOR
TO
INSERTING
ACCESORIES
IN
THE
GUIDES
CHECK
ON
THE
PERFECT
ATTACHMENT
OF
THE
SUPPORT
WITH
THE
MOUNTING
SLOT
ON
THE
COLLIMATOR
(
TOLERANCE
MAX
. ± 0,5
MM
).
A
FAULTY
ATTACHMENT
COULD
BE
DANGEROUS
AND
IT
COULD
CAUSE
THE
FALL
OF
ACCESSORIES
Summary of Contents for R 302F/A
Page 4: ...MTR302 A R302F A R302L A DHHS Contents 2 ...
Page 77: ...info_cl fm ...