General Alarms
14.1. Manual Alarms Test
Under Normal Operation Mode and non-alarm conditions, the buzzer (audible alarm),
vibration, visible alarms, and backlight can all be tested anytime by pressing [Y/+]. If any
alarm does not respond, connect the MicroRAE to a PC running ProRAE Studio II and
check the alarm settings to make sure all alarms are enabled. If any alarms are enabled
but not functional, the instrument should not be used. Contact RAE Systems Technical
Support.
Alarm
Type
Buzzer & LED
Display
Vibrator
Reading
Backlight
Priority
Oxygen
Low Low
3 beeps/second “LOW”
Oxygen
sensor on
display
400ms
-
On
Highest
Low
Alarm
3 beeps/sec
“LOW” at
sensor
location,
ALARM
icon on
400ms
Blinking
reading
On
LEL
failure
3 beeps/second “FAIL” LEL
sensor on
display
400ms
-
On
Confidence
LED flash
Blinking
Confidence
LED for 3
seconds
“OK” on
display
-
-
-
GPS
Offline
-
Blinking
“GPS” icon
-
-
-
Lowest
* For oxygen, “LOW Oxygen” means a concentration is lower than the low alarm limit.