PROGRAMMING AREARAE-Gamma
4-44
4.8.5 Change PID Lamp Type
This programming menu only applies to the monitor, which is
equipped with a PID detector option. There are two different
energy UV lamps available for the PID sensor: 10.6 eV and
11.7 eV. The user must select a lamp in order for the
instrument to calibrate correctly.
1. “Change PID Lamp Type?” is the fourth submenu item.
Press [Y/+], the display shows the current PID lamp
selection: “PID Lamp = 10.6 eV?”
2. Press [Y/+] to accept the currently displayed PID lamp and
automatically goes into Change LEL/VOC Gas Selection.
Press [N/-] to switch to the other lamp. Press [MODE] to
exit this submenu and move to the next submenu.
3. If there is any change to the existing selection, press [Y/+]
and the display will show “Save?” Then, press [Y/+] to
accept the [N/-] to discard and move to the next submenu.
4.8.6 Change RAD Unit
This programming menu allows the user to choose the display
unit for radiation measurement. The default display unit is
µR/h.
1. “Change RAD Unit » is the fifth submenu item. Press [Y/+]
and the current radiation unit selection is: “RAD Unit =
µR/h?”
2.
Press [Y/+] to accept the current selection and go to the
next submenu. Press [N/-] to switch to µSv/h. Press
[MODE] to exit this submenu and return to the first
submenu.
3.
If there is any change to the existing selection, press [Y/+]
and the display will show “Save?” Then, press [Y/+] again
to accept or press [N/-] to discard changes and return to
the first submenu.