PROGRAMMING AREARAE-Gamma
4-23
4.6.4 Enable/Disable Datalog
The user can choose to enable or disable the datalogging
function on each individual sensor. This allows The user to
selectively log certain sensor readings of interest.
1. When the “Enable / Disable Datalog?” appears, press
[Y/+]. The display shows the entire installed sensor in the
monitor. A previously selected sensor to be datalogged is
marked with an “*” next to the sensor name. The cursor is
blinking at the first sensor location. Press [Y/+] to enable
the datalogging for the sensor and [N/-] to disable the
datalogging.
TOX1
VOC
Gamma
CO*
VOC*
Gmma
LEL*
pick
OXY*
LEL
OXY
2. Press [MODE] momentarily to move from one sensor
location to the next one. Repeat Step 2 until all the sensors
need to be datalogged are selected. Press and hold
[MODE] for 1 second to save the new sensor selection.
3. The display shows “Save?” To confirm the new selection,
press [Y/+] to accept the change. Press the [N/-] or
[MODE] to discard the change and return to the first
datalog submenu.