- 39 -
18.4.3. Readout
Menu readout contains data related to WEIGHING PROCESS, such as:
Filter
Each measuring signal is subject to electronic processing
before it is displayed. The filtering is carried out to achieving
correct measurement parameters, specific to a stable signal,
i.e. one that is ready for reading by a balance user.
Balance user is able, in a limited range, to change the signal processing
parameters by selecting one of available FILTER settings.
Available filter modes:
o
very fast
o
fast
o
normal
o
slow
o
very slow
When selecting the filtering level, take into
consideration the actual operating conditions of a
balance. In case of very good operating conditions
set the filter to normal or fast, in case of rough
conditions set the filter to slow or very slow..
Caution:
in case of precision balances 3Y series the recommended range of filter is
between very fast
÷
normal
in case of analytical balances and microbalances 3Y series the recommended
range of filter is between normal
÷
very slow
Value release
This parameter determines displaying the stability marker for a
measurement result. There are 3 available settings of value
release parameter
o
fast
o
fast + reliable / recommended value /
o
reliable
Caution:
The speed of stability marker occurrence depends on applied filter value and applied value
release.