- 34 -
15. COMMUNICATION
Menu “Communication” contains balance settings referring to cases
where the balance is one of IT network components. A user should
set means of balance addressing (i.e. balance address), IP address,
subnet mask and gate address.
The above data should comply with the settings of IT network to which the balance is
connected.
16. UPDATE
Balance update is in other words changing the software of a balance
user – it does not change metrological parameters of a balance. The
balance 3Y series enables updating:
application
printers (cooperating with the balance)
The updating process can be carried out using below interfaces:
Ethernet
USB
RS 232
In case of connecting via Ethernet it is required to connect with a computer network
and establishing balance address in the network.
When updating via USB port, connect to balance’s USB port a data storage device
containing a folder
Update
.
The RS 232 interface – the balance automatically searches for the file with update in
current computer structure.
17. LEVELING
The leveling function is bound to an electronic level. The menu
enables determining time interval for consecutive monitoring of the
level status, and the available settings are:
once
every 15 minutes
continuously
The leveling function can also be disabled, if needed. Means of carrying out leveling
process are described in chapter 5 of this user manual.