- 32 -
11. DISPLAY CALIBRATION
A balance 3Y series comprises a terminal featuring a touch screen
display. The display operates my means of reactions for resistance
changes, which is a reaction to touching a specific display field.
As some software functions and settings require precise touching of a
specific display field, the balance menu contains an option enabling
calibrating the display.
Basically, the display calibration function is to restore the reading accuracy of
resistance changes, as its precision may be relatively low. On enabling the function, a
user should touch 5 points marked on the display.
Note:
If the touch screen display operates correctly, then the display calibration function should not
be used.
12. BALANCE USER SELECTION
In case a balance ensures free access to the menu to all of its
users, then starting any process on the balance does not require
any additional activities.
If each balance user has an individual profile, then before starting
any process on a balance they should select and enter their user
profile.
Entering a user profile requires pressing a display field corresponding to operations on
users, and marking a one of previewed user profiles.