Automatic adjustment
This menu option is used for selecting a factor which determines start of
automatic adjustment process. Accessible options:
None – automatic adjustment disabled
Time – adjustment takes place in time intervals declared in menu
<Automatic adjustment time > (11.6)
Temperature – adjustment is triggered by temperature change only
Both – adjustment activation is triggered both by temperature changes
and time interval changes
CAUTION:
Changing the settings of automatic adjustment is enabled only in balances which are
not subject to conformity assessment (i.e. non-verified balances).
11.6.
Automatic adjustment time
<
Automatic adjustment time > determines time interval in which
automatic internal adjustment of a balance is activated. The time interval is
declared in hours and ranges between 1 and 12 hours.
Setting time interval of automatic internal adjustment:
Select option < Automatic adjustment time >
Using displayed menu select appropriate time interval (given in hours)
which is a time gap elapsing from the last carried out internal automatic
adjustment until activating the following automatic internal adjustment.
CAUTION:
Changing the settings of automatic adjustment is enabled only in balances which are
not subject to conformity assessment (i.e. non-verified balances).
11.7.
Adjustment report printout
Parameter <
Report printout > determines whether or not a report from
automatic internal adjustment should be automatically printed on its
completion.
Setting automatic report printout on adjustment completion. Go to parameter:
< Report printout > and select option <YES>.
11.8.
GLP project
GLP is one of means for maintaining documentation from work in accordance
with adopted quality system. Data selected for printing are printed on each
release of a report from balance adjustment.
Balance operator can use in a GLP report below listed information and signs:
adjustment (adjustment mode)
working mode (name of a working mode)
date, time,
user,
balance type,
balance ID,