- 45 -
no
– report is not printed
YES
– report is printed
.
Remember, that if the parameter is set for
YES
, then a report is generated and
sent automatically.
Content of report from adjustment process depends on settings of GLP
parameters. Any option in the GLP submenu which attribute is YES is included
in a report from adjustment process.
P2 GLP – Good Laboratory Practice
P2.1
uSr
|
P2.2
PrJ
|
P2.3
Ptin
|
YES
P2.4
PdAt
|
YES
P2.5
PuSr
|
YES
P2.6
PPrJ
|
YES
P2.7
PId
|
YES
P2.8
PFr
|
YES
Apart from information set in menu group the report contains:
•
Mass of adjustment weight stored in balance’s memory from last
carried out adjustment;
•
Mass of adjustment weight determined in current adjustment process;
•
Adjustment deviation, i.e. difference between the two mass records.