- 82 -
16.4.1.
Format of data sent on pressing of PRINT key
Depending on setting of
P4.5 PStb
parameter only stable measurement or
instantaneous mass indication will be sent.
CAUTION:
For verified balances printout of instantaneous measurement data is
blocked.
Printout format
1
2
3
4 - 12
13
14 - 16
17
18
st
ab
ilit
y
m
ar
ker
sp
ace
si
gn
m
ass
sp
ace
uni
t
CR
LF
stability marker – [space] if stable
-
[?] if unstable
-
[^] if there is an error of range exceeding on +
-
[v] if there is an error of range exceeding on -
sign
– [space] for positive values or [-] for negative values
mass
– 9 marks alignment to the right
unit
– 3 marks alignment to the left
16.4.2.
Format of data sent as response for commands generated
from a computer
On receipt of a command, balance responses with:
XX_A CR LF – command understood, in progress
XX_I CR LF
– command understood, but currently not available
XX _ ^ CR LF – command understood, but max range is exceeded
XX _ v CR LF – command understood, but min range is exceeded
XX _ E CR LF – error occurred while carrying out the command – time
limit exceeded while waiting for stable measurement result (time
limit is characteristic parameter of balance)
XX
– name of command
And followed by: