28
HR-501 User Manual (Ver.2.3)
④
See the picture drawn in the pads and check the spot to
be attached.
⑤
Attach pads to the surface of patient’s skin precisely.
⑥
Link the pads’ connector to the product as shown in the
below picture.
Adult
Children / Infant
■
■
CAUTION
Despite switched placement of pads, AED works properly.
Despite attachment in reverse, make use of the product
according to the audio instruction. Conduct defibrillation
as soon as possible.
In case the pads are attached to the skin incorrectly,
confirm if adhered side of the pads is no longer dry.
If adhered side of the pads is dry, replace with new
pads.