HR-501 User Manual (Ver.2.3)
IP54
Since the AED is an electronic product, electromagnetic
intervention may affect product performance. At the moment
of product operation, minimize intervention by placing
the AED as far as possible from other electronic devices
that may bring about electromagnetic intervention or turn
off other electronic devices. The devices that may cause
electromagnetic intervention are cell phones, radiophones,
and X-ray equipment.
Warning: Use in No Electromagnetic Intervention Area
Battery and electric pads of HR-501 are designed to be
used from -20
℃
to 60
℃
but, in case of long-term storage,
it is encouraged to keep from 0
℃
~ 40
℃
. In case of electric
pads, there may be problem in product performance if
it conducted long-term use and storage out of range of
0
℃
~ 40
℃
.
Warning: Observation of Temperature Guideline for Stable
Operation
IP54 grade is not capable of blocking direct inflow of
liquid completely. In case of water or liquid inflow inside
the AED, there may be malfunction or failure of the AED.
Warning: Do not have the AED exposed to water or other
liquids directly
Operate in case of occurrence of sudden cardiac arrest
in patients only in order to prevent electric shock accident.
Moreover, use of the AED in a simulator or a manikin
may cause malfunction.
Warning: Do not Attempt to Use for the Purpose of Education,
Training, Experiment, and Demo