HR-501 User Manual (Ver.2.3)
■
■
■
■
5)
Keep the product in a place with good light.
6)
At the moment of storage, confirm the status indication light
periodically if the product operates normally.
CAUTION
911
In case of use around flammable gas, there may be a possibility of explosion hazard.
Since the AED is an electronic product, electromagnetic intervention may affect
product performance. At the moment of product operation, minimize intervention
by placing the AED as far as possible from other electronic devices that may bring
about electromagnetic intervention or turn off other electronic devices. The devices
that may cause electromagnetic intervention are cell phones, radiophones, and X-ray.
The battery included in the product is not rechargeable. There may occur explosion
or fire if the battery is recharged, opened, damaged, or burned. In case of use
of pads or battery not manufactured by RADIAN, the product may not operate
properly. It is strongly recommended to use pads and battery authorized by RADIAN.
In case of product storage, remove packing of electric pads (P-303) and do not
keep the pads connector linked to the product. A weak adhesive will be useless
to a patient in an emergency since the adhesive strength decreases from the moment
of opening the pads.