6
HR-501 User Manual (Ver.2.3)
2. Warning and Caution
Please refer to the following for safe use of HR-501.
Certain contents may be elaborated in detail elsewhere in the manual.
Warning: Do not use if the patient shows any reaction or is
consciousness
HR-501 is a portable defibrillator manufactured for
ventricular fibrillation patients. Before using the product,
it is very necessary to check the patient’s consciousness
and respiratory condition; do not use if the patient is
conscious and reacts.
Warning: The AED’s Basic Setting is Adult Mode.
In adult mode, it delivers 150J of electric shock to the
patient. In case of a child, only up to 50J of electric shock
is set to deliver. If patient is under the age of 7, weighting
below 25kg, set to child mode by using Adult/Child selection
button and place pads solidly to appropriate spot.
Shift Right
Left shift
Adult
Children / Infant
•
•