94
Appendix E
Accu-Gold measurements and Feature comparison
The following are measurements that can be assigned to any cell when creating custom profiles. These
are grouped by the measurement source. Note that not all measurements are available when using the
Nugget or Touch WiFi mode.
USB Cable/
AGT Pro
Nugget/
Touch WiFi
Definition
Digitizer
Air Pressure
yes
yes
The atmospheric corrections compensate for the effects of
variations in the current temperature (T in C) and pressure
(P in kPa) on an unsealed ion chamber and uses the
following equation:
Displayed value = (101.3/P) * ((T + 273.15)/295.15) *
Instrument value
Ave. Pulse Duration
yes
yes
Average duration of pulses.
Duration yes
yes
Exposure duration from start to stop trigger. If pre-pulse
detection was activated, duration is the duration of the main
signal without pre-pulse.
Pre-Pulse Detected
yes
no
True if pre-pulse detection was active and pre-pulse signal
was detected.
Pre-Pulse Duration
yes
no
Duration from detection of pre-pulse start to main signal
start.
Pulse Count
yes
yes
Measurement pulse count.
Pulse Duty Cycle
yes
yes
Pulse duty cycle.
Pulse Frequency
yes
yes
Pulse frequency
Pulse Rep. Period
yes
no
Measurement pulse repetition period.
Total Duration
yes
no
Exposure duration from start to stop trigger, ignores pre-
pulse detection.
Trigger Level
yes
yes
Trigger level used for this measurement.
Trigger Sensor
yes
yes
Sensor that was trigger.
Typical Pulse Count
yes
no
Measurement pulse count ignoring scout pulse.
Multi-Sensor
Anode yes
yes
Tube anode setting used for this measurement.
Average kV
yes
yes
Average kV is derived from the ratios of the outputs of
diodes in the multi-sensor, where these averages use the
entire region of interest.
Biased Average kV
yes
no
The average value of kV in the region of interest above an
adaptive kV threshold.
Calibration Description
yes
yes
The name of the calibration and correction that was used
for making the measurement. e.g. "Diagnostic", "Standard",
"Mo Mo", "Mo Mo GE" etc.
Dose yes
yes
Dose Integral
Dose / mAs
yes
no
Dose / mAs
Dose / Pulse
yes
yes
Dose / Pulse
Summary of Contents for ACCU-GOLD2
Page 1: ...dddddddd ACCU GOLD2 USER GUIDE For use with all Accu Gold digitizers and Accu Gold Touch Pro...
Page 34: ...33...
Page 40: ...39 Daughter designation Extracted Region...
Page 43: ...42...
Page 47: ...46...
Page 59: ...58...
Page 99: ...98 Appendix G Carrying cases Accu Gold Soft Cases Model AGSC...
Page 105: ...104 Test Stands and Positioners ADTS W Accu Dose Test Stand AGTS W Accu Gold Test Stand...