59
Custom anode/filter calibrations for GE
Range of Applications
Anode/Filter Tube
kV
Filt (um)
hvl (mm Al)*
Paddle**
Mo/Al
GE DMR
22-49
1.0 mm
0.42-0.61
2.2 mm
Rh/Al
GE DMR
25-49
1.0 mm
0.42-0.61
2.2 mm
Mo/Mo
GE Pristina
22-32
16-44
0.24 -0.51
2.2 mm
Rh/Ag
GE Pristina
27-40
12-80
0.27 - 0.78
2.2 mm
Rh/Ag
IQST GE Pristina
34
-
-
No Paddle
Mo/Cu
GE Pristina
40-49
250
2.48-3.27
2.2 mm
Rh/Cu
GE Pristina
40-49
250
2.42-3.20
2.2 mm
Mo/Mo
GE Essential
22-48
16-44
0.24-0.51
2.2 mm
Mo/Rh
GE Essential
22-48
6-42
0.20-0.56
2.2 mm
Rh/Rh
GE Essential
25-48
12-80
0.27-0.78
2.2 mm
W/Al
GE Definium (diag)
Same as Diagnostic
No Paddle
W/Al
GE Optima (diag)
Same as Diagnostic
No Paddle
Custom anode/filter calibrations for Philips
Range of Applications
Anode/Filter
Tube
kV
Filt (um)
hvl (mm Al)*
Paddle**
W/Al
Philips Microdose 20-50
300-3030 0.16-1.96
No Paddle
Custom anode/filter calibrations for Siemens
Range of Applications
Anode/Filter Tube
kV
Filt (mm)
hvl (mm Al)*
Paddle**
W/Al
Siemens Straton
Tube CT with W
filter 40-140
0-29
5-10
MDL8155
filter
required
W/Ti
Mammomat
Revelation
41-50
1.0 2.45-3.0 No
Paddle
Custom anode/filter calibrations for Hologic
Range of Applications
Anode/Filter Tube
kV
Filt (um)
hvl (mm Al)*
Paddle**
W/Cu Hologic
Dimensions
38-49 174-325
1.54-3.52
No
Paddle
Summary of Contents for ACCU-GOLD2
Page 1: ...dddddddd ACCU GOLD2 USER GUIDE For use with all Accu Gold digitizers and Accu Gold Touch Pro...
Page 34: ...33...
Page 40: ...39 Daughter designation Extracted Region...
Page 43: ...42...
Page 47: ...46...
Page 59: ...58...
Page 99: ...98 Appendix G Carrying cases Accu Gold Soft Cases Model AGSC...
Page 105: ...104 Test Stands and Positioners ADTS W Accu Dose Test Stand AGTS W Accu Gold Test Stand...