76
DDX6-W (DISCONTINUED)
The DDX6-W is a solid-state dose sensor for diagnostic range measurements. It is a small solid-state
sensor for single exposures and fluoroscopy in the diagnostic range. It measures dose, dose-rate, and
time and is passively corrected for beam quality. It may be used simultaneously with an AGMS-D multi-
sensor in entrance / exit exposure measurements.
Sensor
DDX6-W
Min Rate
2 R/s
20 nGy/s
Max Rate
23 R/s
205 mGy/s
Min Dose
10 R
100 nGy
Max Dose
64 kR
559 Gy
Accuracy
±5% using X-rays @ 80kVp with 2.5 mm Al total filtration (IEC
61267 RQR-6)
Energy Dependence
+5% from 50 kVp to 120 kVp at 2.5 mm Al
Filtration Dependence
+5% to -10% from 2.5 mm Al to 23 mm Al