69
10X6-500 Leakage Measurement Chamber
A single sensor leakage measurement solution for very low-level radiation measurements such as
shielding, leakage, irradiators and environmental that is part of the Accu-Gold family of systems.
Equivalent to the Fluke® 96010A Ion Chamber used in the Fluke® Radiation Leakage Detection System,
this chamber meets 21 CFR 1020.30(k) for leakage measurement requirements.
Chamber
10X6-500
Min Rate
62 nR/s
0.6 nGy/s
Max Rate
130 mR/s
1.2 mGy/s
Min Dose
1 nR
0.01 nGy
Max Dose
45 R
0.4 Gy
Cine Specifications
N/A
Calibration Accuracy
±4% using X-rays @ 100kVp and 3.96 mm AL HVL
Exposure Rate Dependence
<5% up to 5 Gy/hr (570 R/hr)
Energy Dependence
±5%, 40-150 kVp (RQR and RQA Beam Qualities)
Construction
Truncated cylinder, polycarbonate walls and electrode;
conductive graphite exterior coating; 522 cm
3
active volume;
0.54 kg.