Note for Mammography Sensor users.
The Mammographic Sensors, models RAPD-M, AGMS-M, AGMS-DM+ and AGMS-M+ are calibrated
using a 2.2 mm thick polycarbonate compression paddle placed in beam for all measurements (kVp, dose,
hvl). If you are making measurements where it is not practical to use the machine’s paddle in beam, use
this 2.2 mm polycarbonate plate (Model 8154 - Simulated Paddle) on top of the sensor during your
measurements.