40
RIDA qLine® autoBlot | User manual
the sample tube for the instrument to pick up. If inspection results in no issues being
detected, the user should verify that tips of the correct specification are being used.
If the transfer issue states [BarcodeIncorrect], it means the instrument has rotated the
carousel to the position that it found the target sample tube in from the verification run,
but somehow the barcode no longer matches what is expected. In this situation, the user
should open the door and remove the carousel. While the user has access to the carousel,
they should ensure that the barcode points fully outwards and is not smudged or
damaged in any way; a fresh sticker should be placed if damage is found. When done,
the user should put the carousel back inside the instrument and close the door.
Whenever any of the above situations arise, the user is given 4 options as shown in
Figure 34:
1. Abort
Abort the whole work-list. This is the same as aborting from the main processing
screen.
2. Skip
Skips the sample transfer for the current trough; this relies on the user manually
doing the transfer before pressing skip.
3. Skip All
Skips the sample transfer of all remaining troughs; this relies on the user manually
doing the transfers before pressing skip.
4. Retry
Restarts the sample transfer for the current trough (which includes another
3 attempts).