31
Both the autoBlot and the RIDA qLine® Soft must be
connected with the same network. The TCP/IP
communication must be enabled in all parties.
The connection is recommended to be done by an
IT specialist.
During the installation the input mask the IP address,
the port, the name of the device and the serial number
must be entered.
If more than 1 autoBlot should be connected to the
RIDA qLine® Soft the parameters of the other devices
(in maximum 4) must be entered manually into the
qsoft.ini-file. In order to find the qsoft.ini file please
enter %appdata%\R-Biopharm\qsoft into the address
bar of the windows explorer and confirm with the
enter key.
Please open the qsoft.ini file by double click and
go to the section as shown in Fig. 22:
For all connected machines the Description (name), the serial number and the IP address
must be entered.
The software versions of all connected devices must be identical. In case a software update
for the machine is available all connected machines must be updated with the software
update and the software version must be entered into the qsoft.ini file. For that purpose
please look for the section in the qsoft.ini file and enter the software version after the
update. If the software version of the machine and the software version which is entered in
the qsoft.ini is different, the connection of our RIDA qLine® Soft via the R-link software is
not possible.
Fig. 22:
Connected machines