Puma 20/40
Trouble shooting | 187
DX2-ACT Flash Code List
Note:
These flash codes are the Actuator Module flash codes that are visible on LED 1 of the Actuator Module. They are not the DX
System flash codes that are visible on the DX System status LED on the Master Remote. For DX System flash codes, see DX2
faultfinding table
Flash
Code
Fault
source
Sub
code
(HHP)
Meaning
01
User
00
Wizard programming in progress
• Finish programming and then turn the system off and on.
• Make sure that you wait 10 seconds after you have turned the system off before you turn it
on again.
01
Power Down in progress
• Wait 10 seconds before you turn the system on again.
02
Joystick source changed to a joystick that is not in the centre position
• Release the joystick to the centre
03
Invalid Actuator Profile (A-F) or direction demand
• If this happens often, contact the Service Department of Sunrise Medical HCM
02
Slow function
battery
none
see DX2 System condition indications
00
Battery Voltage too low or too high
• Check the batteries and the cables
• Batteries may be empty: charge the batteries
• Batteries may be damaged: replace the batteries
Batteries may be overcharged: if driving downhill, slow down
03
Actuator
00
01
02
03
Internal current or Voltage fault
• Contact the Service Department of Sunrise Medical HCM
04
05
Actuator motor voltage is not what it should be during drive
Possible motor short circuit
• check the actuator cables for damage
• check that the actuator is not faulty
• Otherwise internal controller fault, contact the Service Department of Sunrise Medical HCM
06
Failed To Stop
07
Too Many Fast Current Limit Events
08
Driving Stopped Due To Trip
All
other
Internal fault
• Contact the Service Department of Sunrise Medical HCM
06
Configuration
01
Module ID collision
• Check that multiple Actuator Modules in the system do not have the same ID
02
System check failed
03
Missing system check
05
DX System internal fault
Actuators will not move when the DX System itself is in a fault state
All
other
Internal DX2-ACT configuration fault
• Check DX BUS cables
• Turn the system off and on several times.
Make sure to wait 10 seconds before you turn the system on again.
• If turning the system off and on does not help, contact the Service Department of Sunrise
Medical HCM
Summary of Contents for puma 20
Page 1: ...Puma 20 40 TECHNICAL MANUAL EN Powered wheelchair ...
Page 2: ...Puma 20 40 2 ...
Page 44: ...Puma 20 40 44 Spare parts Cable routing Shark Drive and lights ...
Page 99: ... 99 Puma 20 40 99 5 1 1 2 2 110648 03 034_5spaak_2016 ...
Page 109: ... 109 Puma 20 40 109 3 3 1 2 110648 03 035 4 110648 03 013 5 6 7 8 ...
Page 118: ...118 Puma 20 40 118 8 9 ...
Page 130: ...130 Puma 20 40 130 11 1 2 110648 03 042_5spaak_2016 ...
Page 144: ...144 Puma 20 40 144 11 1 2 110648 03 087 12 3 1 2 110648 03 088 ...
Page 148: ...148 Puma 20 40 148 5 1 2 110648 03 042_5spaak_2016 ...
Page 150: ...150 Puma 20 40 150 5 1 2 110648 03 042_3spaak_2016 ...
Page 152: ...152 Puma 20 40 152 5 6 1 2 110648 03 042_5spaak_2016 ...
Page 160: ...160 Puma 20 40 160 14 3 1 2 110648 03 083 15 110648 03 903 2 2 3 3 1 ...
Page 162: ...162 Puma 20 40 162 5 FW D RW D 6 110648 03 010 6 110648 03 093 ...
Page 163: ... 163 Puma 20 40 163 7 FW D RW D 6 110648 03 094 8 110648 03 903 2 2 3 3 1 ...
Page 168: ...168 Puma 20 40 168 5 10º 110648 03 906 ...
Page 170: ...170 Puma 20 40 170 5 Note If needed fixate the cable with a new tie wrap 6 2 1 110648 03 908 ...
Page 198: ...Puma 20 40 198 ...
Page 199: ...Puma 20 40 199 ...