Puma 20/40
Trouble shooting | 191
Possible cause
Remedy
Joystick Displaced at
Power-up
The most common cause of this trip is if the joystick is deflected away from the center when the
control system is being switched on. When the control system is switched on, the battery gauge
will blink for a short time. Check that the user is not deflecting the joystick before the blink finishes.
If the problem persists, refer to section .37.
Communication Error
The most likely cause of a communication error is a defective cable between the Power Module
and the Joystick Module. The cable should be checked for damage, and replaced if found to
have a fault. The Joystick Cable can only be replaced by a person authorized by the wheelchair
manufacturer. Refer to Section 1.3 for the replacement procedure.
If the problem persists then either the Power Module or the Joystick Module could be defective.
Refer to Section 5.
Inhibit 2 Active
This occurs when the Inhibit 2 Input is active. The Inhibit 2 input is via the INH-2 way connector
and is normally associated with speed limit or actuator functions. The operation of Inhibit 2 will
depend upon the programmed settings and the wheelchair on which it is being used. Check all
wiring and switches connected to Inhibit 2. If these appear to be in working order, then the Power
Module may be defective. Refer to Section 5.
Inhibit 3 Active
This occurs when the Inhibit 3 Input is active. The Inhibit 3 input is via the 3 way on-board charger
(OBC) and is normally associated with this function. The operation of Inhibit 3 will depend upon
the programmed settings and the wheelchair on which it is being used. Check all wiring, switches
and OBC (if fitted) connected to Inhibit 3. If these appear to be in working order, then the Power
Module may be defective. Refer to Section 5.
Actuator Motor Wiring Trip
This occurs when the control system detects a fault in the wiring to either actuator motor. Check
the motor connectors and wiring. If the trip is still present after the above checks have been made,
then the Power Module may be defective. Refer to Section 5.
Current Limit Active
This occurs when the control system operates above the Current Limit Threshold for a periode of
time greater than the Current Limit Time. It has been designed to notify the Healthcare Technician
that the control system has operated outside of its programmed range.
High Temperature
This occurs when the control system reaches its Temperature Threshold and thus becomes to
hot. The controller will not allow drive until the controller to cooled down. An entry is made in the
system log each time the controller gets too hot and goes out of drive.
Summary of Contents for puma 20
Page 1: ...Puma 20 40 TECHNICAL MANUAL EN Powered wheelchair ...
Page 2: ...Puma 20 40 2 ...
Page 44: ...Puma 20 40 44 Spare parts Cable routing Shark Drive and lights ...
Page 99: ... 99 Puma 20 40 99 5 1 1 2 2 110648 03 034_5spaak_2016 ...
Page 109: ... 109 Puma 20 40 109 3 3 1 2 110648 03 035 4 110648 03 013 5 6 7 8 ...
Page 118: ...118 Puma 20 40 118 8 9 ...
Page 130: ...130 Puma 20 40 130 11 1 2 110648 03 042_5spaak_2016 ...
Page 144: ...144 Puma 20 40 144 11 1 2 110648 03 087 12 3 1 2 110648 03 088 ...
Page 148: ...148 Puma 20 40 148 5 1 2 110648 03 042_5spaak_2016 ...
Page 150: ...150 Puma 20 40 150 5 1 2 110648 03 042_3spaak_2016 ...
Page 152: ...152 Puma 20 40 152 5 6 1 2 110648 03 042_5spaak_2016 ...
Page 160: ...160 Puma 20 40 160 14 3 1 2 110648 03 083 15 110648 03 903 2 2 3 3 1 ...
Page 162: ...162 Puma 20 40 162 5 FW D RW D 6 110648 03 010 6 110648 03 093 ...
Page 163: ... 163 Puma 20 40 163 7 FW D RW D 6 110648 03 094 8 110648 03 903 2 2 3 3 1 ...
Page 168: ...168 Puma 20 40 168 5 10º 110648 03 906 ...
Page 170: ...170 Puma 20 40 170 5 Note If needed fixate the cable with a new tie wrap 6 2 1 110648 03 908 ...
Page 198: ...Puma 20 40 198 ...
Page 199: ...Puma 20 40 199 ...