QUICKIE ELECTRONICS SECTION
General Wheelchair Test
Complete the following test to verify Time On, Time
Driven, and Internal Control Box Temperature (Motor
Controller).
Connect the Quickie iQ or HHP to the wheel-
1.
chair.
Turn on the wheelchair
2.
If using the HHP, select Monitor, then System,
3.
then Others. (
fi
g 2.23.30)
If using the Quickie iQ, select Monitor Parameter,
4.
then Motor Controller, then Others. (
fi
g 2.23.31)
Others
+ + -
Time On
3
h
Time Driven
0
h
Int c/box Temp
22
deg
If using the HHP, record Time On, Time Driven,
5.
and Internal Control Box Temperature. If using
Quickie iQ record Time On, Time Driven, Internal
Control Box Temperature, and Motor 1 and Motor
2 Temperature
Refer to the following notes:
6.
Internal Control Box Temperature and Motor
•
Temperate are valuable information for trouble-
shooting Thermal Rollback Issues.
Motor maximum temperatures are approximate-
•
ly 100 degrees C.
Time On and Time Driven are important values
•
to record in case issues are reported after re-
pairs are completed on the wheelchair.
Fig 2.23.30
Fig 2.23.31
Summary of Contents for Groove
Page 1: ......
Page 2: ...2 2006 Sunrise Medical ...
Page 14: ...14 2006 Sunrise Medical INTRODUCTION ...
Page 132: ...RHYTHM SERVICE MANUAL PAGE 0 14 2009 SUNRISE MEDICAL ...
Page 138: ...RHYTHM SERVICE MANUAL 2009 PAGE 2 4 SUNRISE MEDICAL Section 2 ...
Page 144: ...RHYTHM SERVICE MANUAL 2009 PAGE 3 6 SUNRISE MEDICAL ...
Page 148: ...RHYTHM SERVICE MANUAL 2009 PAGE 4 4 SUNRISE MEDICAL ...
Page 162: ...RHYTHM SERVICE MANUAL 2009 PAGE 6 6 SUNRISE MEDICAL ...
Page 168: ...RHYTHM SERVICE MANUAL 2009 PAGE 8 4 SUNRISE MEDICAL Tilt to Interface Mounting Section 8 ...
Page 179: ...RHYTHM SERVICE MANUAL PAGE 9 1 2009 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide ...
Page 180: ...RHYTHM SERVICE MANUAL 2009 PAGE 9 2 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide ...
Page 181: ......
Page 182: ...182 2006 Sunrise Medical ...