POWER BASE SECTION
Output Shaft
Hubs and Bolts
Inspect: Tapered Shaft – Check keyway for wear – check for physical deterioration
Use a 19mm wrench to remove the nut off of end of shaft. Take the washer away. Use a pulley puller to remove the
Hub Assembly from the output shaft. (Fig 1.25&1.26)
Note: Do not lose the hub assembly locating woodruff key.
Rhythm
Groove
Rhythm
Groove
Fig 1.26
Fig 1.25
Fig 1.27
Fig 1.28