SEATING SECTION
Shear Actuator Removal
In order to remove the Shear Actuator, the Gas Strut
and Recline Actuator must be removed. Refer to the
Recline Actuator Removal section in this tech manual.
Once the Recline Actuator is removed, complete
1.
the following steps to remove the shear actuator.
Unplug the shear actuator by disconnecting the
2.
6-pin connector (Fig 3.125).
Using a 7/16” open-end wrench and a 3/16” hex
3.
wrench, remove the top and bottom mounts.
Lift the actuator out of the back shroud by pulling
4.
the top mount out towards the back of the chair,
and then remove the Shear Actuator from of the
bottom mount.
Reverse the process to reinstall a new Shear Ac-
5.
tuator. The part number is 012038 (Fig 3.126).
Fig 3.125
Fig 3.126
Shear Actuator Removal
Summary of Contents for Groove
Page 1: ......
Page 2: ...2 2006 Sunrise Medical ...
Page 14: ...14 2006 Sunrise Medical INTRODUCTION ...
Page 132: ...RHYTHM SERVICE MANUAL PAGE 0 14 2009 SUNRISE MEDICAL ...
Page 138: ...RHYTHM SERVICE MANUAL 2009 PAGE 2 4 SUNRISE MEDICAL Section 2 ...
Page 144: ...RHYTHM SERVICE MANUAL 2009 PAGE 3 6 SUNRISE MEDICAL ...
Page 148: ...RHYTHM SERVICE MANUAL 2009 PAGE 4 4 SUNRISE MEDICAL ...
Page 162: ...RHYTHM SERVICE MANUAL 2009 PAGE 6 6 SUNRISE MEDICAL ...
Page 168: ...RHYTHM SERVICE MANUAL 2009 PAGE 8 4 SUNRISE MEDICAL Tilt to Interface Mounting Section 8 ...
Page 179: ...RHYTHM SERVICE MANUAL PAGE 9 1 2009 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide ...
Page 180: ...RHYTHM SERVICE MANUAL 2009 PAGE 9 2 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide ...
Page 181: ......
Page 182: ...182 2006 Sunrise Medical ...