26
Camera soft error state indication
Flash Red: 1
Flash Yellow:1
The camera is over-temperature. Camera has exceeded the 40°C maximum recommended
operating temperature for the internal electronics and enclosure. This sequence will repeat
every four seconds as long as the camera remains “over- temperature”.
Flash Red: 1
Flash Yellow: 2
The CCD is under-temperature. The CCD has exceeded the -40°C minimum recommended
operating temperature. This sequence will repeat every four seconds as long as the CCD
remains “under-temperature”.
Flash Red: 1
Flash Yellow: 3
The camera DC supply voltage is below the 11V minimum recommended operating level. This
sequence will repeat every four seconds until the supply voltage is raised above this threshold.
Flash Red: 1
Flash Yellow: 4
The camera DC supply voltage is above the 14V maximum recommended operating level. This
sequence will repeat every four seconds until the supply voltage is lowered below this threshold.
Flash Red: 1
Flash Yellow: 5
The filter wheel has encountered a problem and is no longer operational. The camera will still
function, but will no longer attempt to move the filter wheel. The filter wheel may be in any
position. Power to the camera must be turned off and on to reinitialize the filter wheel.
Flash Red: 1
Flash Yellow: 6
The shutter has encountered a problem and is no longer operational. The camera will still
function, but will no longer attempt to operate the shutter. The shutter may be in any state.
Power to the camera must be turned off and on to reinitialize the shutter.