Rotor-Gene Q MDx CE User Manual 02/2022
28
USB and RS-232 serial cable
International power cable set
PCR Tubes, 0.2 ml (1000)
Strip Tubes and Caps, 0.1 ml (1000)
Once all these components have been removed from the box, remove the foam packing on top of
the Rotor-Gene Q MDx. Carefully remove the Rotor-Gene Q MDx from the box and unwrap the
plastic cover. Open the lid by sliding it towards the back to access the reaction chamber.
The following items are already installed inside the Rotor-Gene Q MDx:
72-Well Rotor (this rotor is blue in color)
72-Well Rotor Locking Ring
A laptop computer may be included in the packaging, depending on your order details.
4.1.2
Hardware installation
Once the Rotor-Gene Q MDx has been unpacked, proceed with installation as described below.
CAUTION
Damage to the instrument
When the Rotor-Gene Q MDx is started immediately after delivery in cold
climates mechanical parts can block. Allow the instrument to acclimatize to
room temperature for at least an hour before turning the instrument on.
Proceed as follows:
1.
Place the Rotor-Gene Q MDx on a level surface.
2.
Ensure that there is sufficient space behind the instrument for the lid to open fully.
3.
Ensure that the power switch at the back of the instrument can be reached easily.
4.
Do not obstruct the back of the instrument. Ensure that the power cord can be easily detached
if required, to disconnect power to the instrument.
5.
Connect the USB cable or RS-232 serial cable supplied to a USB or communications port on
the back of the computer.
6.
Connect the USB or RS-232 serial cable to the back of the Rotor-Gene Q MDx.