CAUTION
Risk of personal injury and material damage
Do not try to open the lid during an experiment, or while the Rotor-Gene Q MDx
is spinning. Otherwise, if you overcome the lid lock and reach inside, you risk
contact with parts that are hot, electrically live, or moving at high speed, and
you may injure yourself and damage the instrument.
CAUTION
Risk of personal injury and material damage
If you need to stop an experiment quickly, turn off the power to the instrument,
then open the lid. Let the chamber cool before reaching inside. Otherwise you
risk injury by touching parts that are hot.
CAUTION
Risk of personal injury and material damage
If the equipment is used in a manner not specified by the manufacturer, the
protection provided by the equipment may be impaired.
CAUTION
Risk of personal injury and material damage
Loose paper underneath the Rotor-Gene Q MDx interferes with instrument
cooling. It is recommended that the area beneath the instrument is kept free of
clutter.
CAUTION
Damage to the instrument
Always use a locking ring on the rotor. This stops caps from coming off tubes
during an experiment. If caps come off during an experiment, they may damage
the chamber.