13 Troubleshooting and Problem Solving
This chapter is intended for users and programmers. The
descriptions shall help users to solve problems and correct errors
in normal operation of the .
Error:
No response to MODBUS requests.
Cause:
MODBUS commands are being incorrectly sent.
Correction:
Implement the MODBUS-Protocols according to
standard. See example in section 10.7.
Error:
No response to MODBUS requests.
Cause:
Incorrect master interface settings.
Correction:
Set the master serial interface settings to 57600, 8, 1,
n. See test example in section 10.7.
Error:
No response to MODBUS requests.
Cause:
Incorrect MODBUS Address is being used.
Correction:
Try communication with MODBUS address „0“.
Test example in section 10.7.
Error:
No response to MODBUS requests.
Cause:
Checksum has been incorrectly calculated
Correction:
Calculate Checksum according to LRC (See section
12.1). See test example from section 13.
Error:
No power input.
Cause:
Power supply not connected.
Correction:
Check pin assignment in section 3, test and make any
necessary alterations.
Error:
No power input.
Cause:
External power source has not been switched on.
Correction:
Switch on power source.
Error:
No Communication
Cause:
Reset-wire (see section 3) is grounded.
Correction:
Examine the level of the reset wire and, if necessary,
connect via a resistor to +5V.