FLUO SENS integrated Manual 

January 2013 

 

10.5 Register Assignment 

 

For communication purposes, the MODBUS-ASCII Protocol has 
been installed in the . The following sections describe the storage 
allocation.  

 

10.5.1 Definitions 

 
 

D1 = Measurement Values from Detection Channel 1 
D2 = Measurement Values from Detection Channel 2 
E1 = Excitation Channel 1 
E2 = Excitation Channel 2 

Cycles 

Number  of  measurements  to  be  carried  out  after  the  Start 
command has been sent.  

Cycle 
Time 

Interval in seconds between two measurements within a cycle. 

Start 
mode 

 

 

By command

:   If a „1“ has been written in location 512, 

the measurement will be started  

 

Trigger  single

:  Measurement  is  started  by  sending  the 

“Start  Method  Command”.  One  data  sequence  is  taken 
for each h/w trigger until the method is finished. 

 

Trigger  measurement

:  Measurement  is  started  by 

sending  the  “Start  Method  Command”,  all  values  are 
taken after one h/w trigger 

 

Auto  command

:  Measurement  is  started  at  power  on, 

the stored method will run once 

 

Auto  trigger  single

:  Measurement  is  started  at  power 

on,  one  value  is  taken  for  every  h/w  trigger  until  the 
method is finished 

 

Auto trigger measurement

:  Measurement is started at 

power on, all values are taken after one h/w trigger 

 

 

 

 

Method Type 

 

E1D2: Emission on channel 1, detection on channel 2 
S_E1D1: Scope mode E1D1. The detector records  
1500 data points and saves them in registers 513-3513.  

Dark Signal 
Type 

 

No dark signal (LED off):No dark signals recorded 
Dark  signal  once:  At  the  beginning  of  a  measurement 
cycle, 

one

 dark signal will be recorded. 

Dark  signal  every:  At  the  beginning  of  a measurement 
cycle,  dark  signal  will  be  recorded 

for  every 

measurement value

Average 

Number  of  measurement  values  used  to  calculate  an 
average. 

LED Mode 

Toggle: LED controlled by the detector  
Manual:  LED  must  be  set  manually  before 
measurement is starting 

Trigger Delay 

Wait  time  in  ms  before  a  measurement  is  started  via 
trigger impulse 

On Delay LEDx 

The  detector  waits  for  “On  Delay  LEDx”  ms  after  the 
LEDx  has  been  turned  on  before  a  measurement  is 
carried out. 

Off Delay LEDx 

 

The  detector  waits  for  “On  Delay  LEDx”  ms  after  the 
LEDx  has  been  turned  off  before  a  dark  signal 
measurement is carried out. 

LEDx Power  
Default 

“LEDx ” standard LED set value for LED x 
 

LEDx Power Max 

Maximum allowed set value for “LEDx Power” 

LEDx Power Min 

Minimum allowed set value for “LEDx Power” 

Summary of Contents for FLUO SENS

Page 1: ......

Page 2: ...Manual...

Page 3: ...s Trademarks and labels used are the property of their respective owners Copyright Information User Manual for FLUO SENS integrated Document Number ESMO30 DH 1001 Document Revision 05 January 2013 Cop...

Page 4: ...8 3 Switching off the Device 12 9 Operation 13 9 1 MODBUS ASCII 13 10 Configuring a MODBUS Protocol 14 10 1 Longitudinal Redundancy Check LRC 14 10 2 MODBUS Function Codes 14 10 3 Function Code 03 Re...

Page 5: ...tors or for online measurements in automated processes The confocal optics allows contact free measurements on surfaces and in liquids in the presence of ambient light The highly sensitive detector me...

Page 6: ...ion wavelengths 265nm 980nm Available detection wavelengths 365nm 980nm Specifications given in the list above are for reference only Exact specifications for the individual detectors are given in the...

Page 7: ...pe 150mA The FLUO SENS integrated features a 10way flex cable in the rear part of the housing see 4 1 and 4 2 This flex cable is used for communication and power supply Dimensions are shown left Flex...

Page 8: ...etector is connected to ground return To achieve precise alignment between the detector and the sample the front lens is designed with a 8f8 fit shown bottom left ESMO40 MB xxxx Detector Dimensions 4...

Page 9: ...5 C Relative Humidity 20 to 70 rel humidity 5 1 Environment If you plan to use the device in a working environment prone to dust and dirt you will need to clean the device regularly For cleaning use a...

Page 10: ...ay heat up in normal operating mode Tip Never touch the detector s optics Danger Do not look directly into the optics Detectors emit strong light radiation that can cause permanent damage to the eye D...

Page 11: ...d 6 1 Start up The is ready for use upon delivery 6 2 Mounting The housing possesses two M3 threads on both long sides for mounting purposes The positions of the threads are given in the drawings in c...

Page 12: ...devices however the built in drivers limit the number to 64 for RS485 and 16 for TTL versions The MODBUS address for each device must be set at a value between 1 and 254 before being connected Each a...

Page 13: ...regulations In Germany this includes the VDE and Employer s Liability Insurance Association regulations as well as the stipulations of the EMC directives 8 1 EMC Safety Precautions In the case of inc...

Page 14: ...ntroller The MODBUS Controllers in the FLUO SENS integrated are slave devices There are no bus conflicts since only a single node sends at a time The master requests communication This request can be...

Page 15: ...the transmitting device which appends the LRC to the message The receiving device recalculates an LRC during receipt of the message and compares the calculated value to the actual value it received in...

Page 16: ...EB Value Hi Register 235 00 Number of Registers Hi 00 Value Lo Register 235 00 Number of Registers Lo 02 Value Hi Register 236 00 LRC 07 LRC F1 End Frame CRLF End Frame CRLF 10 4 Function Code 06 Wri...

Page 17: ...h w trigger until the method is finished Auto trigger measurement Measurement is started at power on all values are taken after one h w trigger Method Type E1D2 Emission on channel 1 detection on chan...

Page 18: ...Page 17 of 24 FLUO SENS integrated Manual January 2013 10 6 FLUO SENS integrated Memory Allocation...

Page 19: ...ark For byte parameters in bytes 0 or 2 like Modbus Adress the given values written to the detector must be multiplied by 256 Example Set Modbus Address to 2 you have to write 2 256 512 to the detecto...

Page 20: ...t value On Value 1 Request 000301040002f6 Response 000304001C5A364D Integer Interpretation 001C5A36h 1858102 counts Conversion counts mV counts 0 000298026 1858102 Counts 553 76mV 10 12 Setting Start...

Page 21: ...nication via Hyper Terminal 12 The FL Digital Program The FL Digital program allows the user to collect and display data from the FLUO SENS integrated Instrument sensitivity parameters can be setup an...

Page 22: ...t example in section 10 7 Error No response to MODBUS requests Cause Incorrect MODBUS Address is being used Correction Try communication with MODBUS address 0 Test example in section 10 7 Error No res...

Page 23: ...s to ensure secure contact Tip If the is operated in high temperature environments the measurement values can deviate somewhat from the real values In this case calibrate the device beforehand under t...

Page 24: ...rtner has the option to demand a reduction in price or withdraw from the contract The same shall apply if replacement of the product fails or cannot be reasonably expected from QIAGEN Lake Constance G...