General Operation
DML 3000 User Manual
6-1
6
General Operation
The DML 3000 measures light in the visible spectrum (300–650 nm) and measures
chemiluminescence from opaque microplates. It is optimally designed for the
measurement of glow-type luminescence.
To measure a microplate using the
digene
HC2 System Software, the following is
required:
The DML 3000 is initialized and has established communication with the HC2
System PC.
A plate layout exists in the
digene
HC2 System Software indicating the microplate
wells to be measured. Refer to digene
HC2 System Software User Manual
for
additional information.
The microplate to be measured has been inserted into the DML 3000.
A mechanical test has been performed.
6.1
Moving the transport locks to the home position
The transport locks must be located in the home position to operate the DML 3000.
Use this procedure to move into the transport locks to the home position.
Open the access door.
Loosen the red thumb screws of both transport locks.
Slide the transport lock on the optical head guide rail all the way to the left and
tighten the thumb screw.
Summary of Contents for DML 3000
Page 5: ...Contents DML 3000 User Manual Contents 3 FCC Declaration D 1 Index Index 1 ...
Page 6: ...Contents Contents 4 DML 3000 User Manual This page intentionally left blank ...
Page 34: ...Functional Description 3 20 DML 3000 User Manual This page intentionally left blank ...
Page 42: ...Installation 5 4 DML 3000 User Manual Example of the DML 3000 with the HC2 System PC ...
Page 43: ...Installation DML 3000 User Manual 5 5 Diagram of connecting a DML 3000 with a RCS ...
Page 44: ...Installation 5 6 DML 3000 User Manual Diagram of connecting a DML 3000 with a LIS ...
Page 45: ...Installation DML 3000 User Manual 5 7 Diagram of connecting a DML 3000 with a RCS and a LIS ...
Page 46: ...Installation 5 8 DML 3000 User Manual This page intentionally left blank ...
Page 54: ...General Operation 6 8 DML 3000 User Manual This page intentionally left blank ...
Page 57: ...Maintenance DML 3000 User Manual 7 3 Example of wiping the interior of the plate mask ...
Page 64: ...Maintenance 7 10 DML 3000 User Manual This page intentionally left blank ...
Page 72: ...Appendix C C 2 DML 3000 User Manual This page intentionally left blank ...
Page 74: ...Appendix D D 2 DML 3000 User Manual This page intentionally left blank ...
Page 76: ...Index Index 2 DML 3000 User Manual Waste electrical and electronic equipment C 1 ...
Page 77: ...Appendix D DML 3000 User Manual 3 This page intentionally left blank ...
Page 78: ...Index Index 4 DML 3000 User Manual This page intentionally left blank ...
Page 79: ...Appendix D DML 3000 User Manual 5 This page intentionally left blank ...
Page 80: ...1058536CS Rev 01 04 2012 Sample Assay Technologies ...