Maintenance
DML 3000 User Manual
7-9
Once dry, align the optical mask with the optical head and gently guide the
optical mask into the optical head.
A light click will be heard when the optical mask is correctly positioned in the
optical head.
Once dry, position the optical mask retainer just above the pins on the optical
head. Push the retainer to the right to snap the optical mask retainer securely to
the right pin of the optical head. Then push down on the left side of the optical
mask retainer to snap it onto the left pin.
Moisten a low-lint paper towel with 70% isopropyl alcohol, and thoroughly wipe
the black metal shelf to the right of the centered optical head. Allow the black
metal shelf to air dry completely.
Moisten a low-lint paper towel with 70% isopropyl alcohol, and wipe the two
steel guide rails and the interior floor. Allow the guide rails and the floor to air dry
completely.
Moisten a low-lint paper towel with 70% isopropyl alcohol, and wipe the inside of
the access door and all other visible interior surfaces not previously cleaned.
Allow to air dry completely.
Gently push the optical head all the way to the right to return the optical head to
the home position.
Failure to move the optical head to the home position will not harm the
instrument, but will cause a loud grinding noise upon powering ON the
instrument. The optical head will move to the home position upon powering ON
the instrument.
Connect the power supply.
The instrument may now be powered ON.
7.2
Decontaminating the DML 3000
To decontaminate the DML 3000, clean the interior of the instrument. See “Cleaning
the interior of the DML 3000,” page 7-1, for additional instructions. Once the cleaning
is complete, wait at least 10 minutes before wiping the interior of the instrument with
a low-lint paper towel moistened with deionized or distilled water.
Summary of Contents for DML 3000
Page 5: ...Contents DML 3000 User Manual Contents 3 FCC Declaration D 1 Index Index 1 ...
Page 6: ...Contents Contents 4 DML 3000 User Manual This page intentionally left blank ...
Page 34: ...Functional Description 3 20 DML 3000 User Manual This page intentionally left blank ...
Page 42: ...Installation 5 4 DML 3000 User Manual Example of the DML 3000 with the HC2 System PC ...
Page 43: ...Installation DML 3000 User Manual 5 5 Diagram of connecting a DML 3000 with a RCS ...
Page 44: ...Installation 5 6 DML 3000 User Manual Diagram of connecting a DML 3000 with a LIS ...
Page 45: ...Installation DML 3000 User Manual 5 7 Diagram of connecting a DML 3000 with a RCS and a LIS ...
Page 46: ...Installation 5 8 DML 3000 User Manual This page intentionally left blank ...
Page 54: ...General Operation 6 8 DML 3000 User Manual This page intentionally left blank ...
Page 57: ...Maintenance DML 3000 User Manual 7 3 Example of wiping the interior of the plate mask ...
Page 64: ...Maintenance 7 10 DML 3000 User Manual This page intentionally left blank ...
Page 72: ...Appendix C C 2 DML 3000 User Manual This page intentionally left blank ...
Page 74: ...Appendix D D 2 DML 3000 User Manual This page intentionally left blank ...
Page 76: ...Index Index 2 DML 3000 User Manual Waste electrical and electronic equipment C 1 ...
Page 77: ...Appendix D DML 3000 User Manual 3 This page intentionally left blank ...
Page 78: ...Index Index 4 DML 3000 User Manual This page intentionally left blank ...
Page 79: ...Appendix D DML 3000 User Manual 5 This page intentionally left blank ...
Page 80: ...1058536CS Rev 01 04 2012 Sample Assay Technologies ...