Operating Procedures
BLU-V User System Manual
09/2016
31
When the start button is pressed, the entered time disappears and the count
starts at zero.
Note:
The setting of the last completed run is saved as the new default.
5.1.2 Setting the light intensity
Press the light intensity button to set the internal LED light intensity.
The default light intensity is 100%.
The light intensity setting can be changed by pushing the light intensity
button until the required value is reached. The increments are 10% each.
The light intensity button is a cyclic push button — after 100%, it begins
again with 10%.
Adjust the light intensity settings based on the recommendations given in
the corresponding BLU-V Kit handbooks. For standard illumination settings
keep the light power at 100%.
Note:
The setting of the last completed run is saved as the new default.
5.1.3 Using the protective shield
Close the protective shield before staring the run.
Keep the protective shield closed until the run is completed.
If the protective shield is opened during a run, the instrument will interrupt
automatically and the photo-activation LEDs will switch off. The time will also
stop. The run and the time will only continue when the protective shield is closed
again.
5.2 Performing the run
1.
To start the run, press the start button. The color of the indicator LED will
change from green to red. The time display will show the elapsed time from
the beginning of the operation in minutes.