EZ2 Connect MDx User Manual 05/2022
147
Figure 146. Protocol run completed screen.
3.
Put on gloves; two pairs are recommended as the piercing unit is sharp.
4.
To prepare the piercing unit, tap
Move down
. The instrument lowers the piercing unit of the
pipettor head.
5.
Open the hood.
6.
Carefully wipe the piercing unit using a lint-free tissue moistened with 70% ethanol. Enclose
individual piercing spike with the tissue, apply firm pressure and twist several times. Repeat
for all piercing spikes.
Figure 147. Cleaning the EZ2 piercing unit.