82
ERROR
DESCRIPTION
ACTION TO CARRY OUT
EFFECT
SOLUTION
The weighing scale
switches off, when the
alarm starts to ring.
Disconnect the 2-pin connector
of the alarm from the cable
controlling it.
The scale works
properly.
The alarm doesn't work properly and must
be replaced.
The scale still
switches off.
The cause could lie in the cable controlling
the alarm. Carry out the action described
in the next point:
Disconnect the cable controlling
the alarm of the scale, by means
of the 7-pin connector.
The scale works
properly.
The cable controlling the alarm is
damaged. Replace it.
The scale still
switches off.
Contact service.
The weight shown by
the scale is not
correct.
Carry out a new calibration of the
system as indicated in the
specific documentation.
The weight shown is
correct.
The calibration of the system was
successful.
The weight shown is
not correct.
Contact service.
The scale does not
switch on, or the
message Low Bat
appears.
Check if the power supply
complies with the technical
features of the product.
The scale switches
on.
The scale does not
switch on: go on to
the next point.
Check if the power supply cable
is interrupted or damaged, and if
the power plug and the battery
clamps are oxydated. If
necessary, replace them.
The scale switches
on.
The scale does not
switch on.
Contact service.