S8000 -100 User Manual
Michell Instruments
47
APPENDIX B
Appendix B
Modbus Register Map
Register Types
Type
Description
uint16
unsigned 16 bit value
uint32
unsigned 32 bit value over two registers
Register names ending with _MS contain the upper 16 bits
Register names ending with _LS contain the lower 16 bits
flags
unsigned 16 bit value where each bit represents a flag, value can be a combination of flags
float
IEEE754 binary32 compatible floating point number
Register names ending with _MS contain the upper 16 bits including the sign and exponent
Register names ending with _LS contain the lower 16 bits
boolean unsigned 16 bit value with only two valid values, 0 = off/disabled/no, 1 = on/enabled/yes
Address Access
Data Type Register Map Definition
Max.
Min.
Default
Instrument Info
0
R|W|F
uint16
MODBUS_ADDRESS
255
1
1
1
R
uint16
INSTRUMENT_ID
42251
42251
42251
2
R
uint32
INSTRUMENT_SERIAL_MS
4294967295
0
0
3
R
INSTRUMENT _SERIAL_LS
4
R
uint16
INSTRUMENT_FIRMWARE_VERSION
65535
0
Version * 1000 (1012 = 1.012)
5
R
uint16
REGISTER_MAP_VERSION
65535
0
Version * 1000 (1012 = 1.012)
Measured and Calculated Values
6
R
float
DEWPOINT_MS [°C/°F]
1000
-1000
N/A
7
R
DEWPOINT_LS [°C/°F]
8
R
float
AMBIENT_TEMP_MS [°C/°F]
1000
-1000
N/A
9
R
AMBIENT_TEMP_LS [°C/°F]
10
R
float
PRESSURE_MS [P]
1000
-1000
N/A
11
R
PRESSURE_LS [P]
12
R
float
RH_MS
100
0
N/A
13
R
RH_LS
14
R
float
PPMV_MS
999999.9
0
N/A
15
R
PPMV_LS
16
R
float
PPMW_MS
999999.9
0
N/A
17
R
PPMW_LS
18
R
float
ABSOLUTE_HUMIDITY_MS
999999.9
0
N/A
19
R
ABSOLUTE_HUMIDITY_LS
20
R
float
MIXING_RATIO_MS
2
0
N/A
21
R
MIXING_RATIO_LS
22
R
float
WETBULB_MS [°C/°F]
1000
-1000
N/A
23
R
WETBULB_LS [°C/°F]
24
R
float
WVP_MS
1000
-1000
N/A
25
R
WVP_LS
26
R
float
FLOW_MS
2000
0
N/A
27
R
FLOW_LS
28
R
float
PERCENT_VOLUME_MS
100
0
N/A
29
R
PERCENT_VOLUME_LS
36
R
uint16
TEMPERATURE_UNIT
1
0
0
0 = °C
1 = °F
Summary of Contents for Michell Instruments S8000 -100
Page 1: ...S8000 100 Chilled Mirror Hygrometer User Manual 97606 Issue 1 June 2021 nbn Austria GmbH ...
Page 51: ...S8000 100 User Manual Michell Instruments 43 APPENDIX A Appendix A Technical Specifications ...
Page 54: ...S8000 100 User Manual 46 97606 Issue 1 June 2021 APPENDIX B Appendix B Modbus Register Map ...
Page 67: ...S8000 100 User Manual Michell Instruments 59 Notes ...