S8000 -100 User Manual
Michell Instruments
29
OPERATION
3.5.7 DCC
Figure 16
DCC Screen
Parameter
Description
Type
DCC heating temperature can either be relative to last measured dew
point or an absolute temperature. Actual temperature or Δ is defined
by ‘Setpoint’.
Available Input:
Relative, Absolute
Setpoint
Mirror heating temperature during DCC, either absolute or relative to
last measured dew point. See ‘Type’ option above.
Available Input:
1...120 °C
Mode
DCCs can either be triggered automatically at every interval or they can
be manually triggered only.
Available Input:
Manual, Auto
Interval
Time between automatic DCCs
Input format:
hh:mm
Limits:
01:00…99:00
Period
Duration of the DCC
Input format:
hh:mm
Limits:
00:01…00:59
Output Hold
Minimum time to hold analog outputs after finishing a DCC
Input format:
hh:mm
Limits:
00:04…00:59
FAST
Turns frost assurance on or off. See Section 3.4.7 for further information
Available Input:
On, Off
FAST SP
Passing this mirror temperature will trigger the frost assurance function
without a DCC
Available Input:
-28...-3 °C
Table 7
DCC Parameters