S8000 -100 User Manual
Michell Instruments
5
INSTALLATION
2.4
Operating Requirements
2.4.1 Environmental Requirements
It is important to operate the S8000 -100 within the following environmental conditions:
Minimum Operating Temperature
5°C
Maximum Operating Temperature
30°C
Maximum Relative Humidity
80%
2.4.2 Electrical Requirements
The S8000 -100 requires the following electrical supply:
•
85...264 V AC, 47/63 Hz, 185 VA max.
•
Alarm outputs comprise two sets of changeover relay contacts, one set
for a
PROCESS
alarm and one set for an
INSTRUMENT FAULT
. Both sets
of contacts are rated at 30 V, 1A. NOTE: THIS RATING MUST NOT BE
EXCEEDED.