3.5.4 Cooler Setup
The Cooler Setup screen is accessed by touching
Sensor Temp readout
on the Main
Screen. Refer to Section 3.4
for detailed information on the operation of the sensor
cooling system.
System Temperature:
OFF
Input out of range:
OFF
System Shutdown:
OFF
Figure 12
Cooler Setup Screen
Parameter
Description
Set-point
Controls the sensor temperature
Limits:
-80…+20
Mode
Changes between Automatic and Manual cooler control
Table 5
Cooler Setup Parameters
In manual mode, the cooler set-point must be maintained higher than the dew point of
the applied gas. A margin of at least 10 °C (18 °F) is recommended.
Cooler Alarm
Warnings
Description
System
Temperature
Cooler heat-sink close to maximum safe temperature.
The environmental temperature may be too hot, or the fan may have
stopped operating.
Continuing to operate the S8000 -100 without addressing this problem
may cause the cooler to overheat.
Input out of
range
Hardware fault
Contact Michell Instruments' service department.
System
Shutdown
Cooler has been automatically disabled to prevent damage.
May be caused by overheating, power supply problem or other safety
issue.
Table 6
Cooler Alarm Warnings
Summary of Contents for Michell Instruments S8000 -100
Page 1: ...S8000 100 Chilled Mirror Hygrometer User Manual 97606 Issue 1 June 2021 nbn Austria GmbH ...
Page 51: ...S8000 100 User Manual Michell Instruments 43 APPENDIX A Appendix A Technical Specifications ...
Page 54: ...S8000 100 User Manual 46 97606 Issue 1 June 2021 APPENDIX B Appendix B Modbus Register Map ...
Page 67: ...S8000 100 User Manual Michell Instruments 59 Notes ...