S8000 -100 User Manual
10
97606 Issue 1, June 2021
INSTALLATION
2.4.5 Analog Output Connections
The two analog outputs can be configured to represent any of the directly measured
or calculated output parameters. They are provided as 2-wire signals from a 6-way
connector located on the rear panel of the instrument
.
Each of these outputs provide a current loop signal (4...20 mA or 0...20 mA). The
0/4...20 mA outputs are active (sourcing) and must be connected to a passive (sinking)
input on the receiving equipment. The configuration of these outputs, i.e. parameter
represented, output type (current loop or voltage) and upper/lower span levels are set
up via the Setup Menu Screen (refer to Section 3.5.5).
These signals may be used to control external systems. During a
DCC
cycle, and for
the hold period following a
DCC
cycle, they are held at the level that they were at
immediately prior to the start of the cycle. When the dew-point measurement is stable,
or if the maximum hold period has expired, they are released and will track the selected
parameter throughout the measurement cycle.
The default settings of these analog outputs are:
Channel 1
:
Dew point, -100...+20ºC
Channel 2
: ppm
V
, 0...3000
NOTE: The analog outputs are only active during the
MEASURE
phase. They
will, therefore, be off after switch-on and remain off until the system enters
the
MEASURE
phase.
The two analog output port connections are made via a single 4-way push fit connector
block. All outputs are 2-wire, positive-going signals referenced to a common 0 V line. To
differentiate between the outputs it is recommended that a black lead be used for each
of the COM (common) lines and a separate color for each of the positive lines.
Figure 6
Alarm and Analog Output Connection
Summary of Contents for Michell Instruments S8000 -100
Page 1: ...S8000 100 Chilled Mirror Hygrometer User Manual 97606 Issue 1 June 2021 nbn Austria GmbH ...
Page 51: ...S8000 100 User Manual Michell Instruments 43 APPENDIX A Appendix A Technical Specifications ...
Page 54: ...S8000 100 User Manual 46 97606 Issue 1 June 2021 APPENDIX B Appendix B Modbus Register Map ...
Page 67: ...S8000 100 User Manual Michell Instruments 59 Notes ...